| Scan Mode |
Fully automatic electronic real-time scanning of linear and convex arrays. |
| Probe Channels |
Channel 96, Channel 128 |
| Number of synchronous probe channels |
Transmission: Channel 64 Reception: Channel 32 |
| Operation mode |
B mode, B/B mode, B/M mode, B/Z mode, M mode: Scan |
| Scope |
3.5MHz 5MHz or more |
| Focus mode |
Transmit: four focus platforms, receive: dynamic focus |
| prescribed ultrasonic frequency |
1.5,2.0,2.8,3.0,3.5,5.0,7.0,7.5,9.0,10.0,12.5MHz |
| Display |
10.4 inch LCD monitor |
| Sound power regulation |
20-95% (10% regulation) |
| Grayscale |
256 |
| Image Display |
Up/Down, Left/Right, Positive/Negative |
| Image Adjustment |
B-增益 : 36~100dB (1dB Step : Adjustable) M-增益 : 36~100dB (1dB Step : Adjustable) B- dynamic range: 35 ~ 95dB (1dB Step : Adjustable) M-Dynamic Range: 35 ~ 95dB (1dB Step : Adjustable) STC : 8-Level Slide Control gamma curve Sharpening function: off, 1-4 adjustable Image edge boost: off, 1-3 adjustable |
| M-Mode Speed Conversion |
1/1, 1/2, 1/4, 1/8, 1/16 B/M-Mode: (1/1) is 1 second per frame. M-Mode : (1/1) is 2 seconds per frame. |
| Measurement method |
Distance measurement: 8 times, area and perimeter measurement: 4 times, volume measurement: 2 times |
| Gestational Age Table/Curve Display |
ESD, ECD, CTD, CHL, CHD, CCRL, STD, SHD, dgsd, dcrl, DHD, DBD, CHD, CBD |
| Data storage |
100 sheets |
| Character Display |
Date, time, (40 characters), patient information (ID: 32 characters, age), probe selection, focus information, scan depth. |
| video signal |
Video signal (PAL or NTSC) |
| External I/O |
Video Output (2 channels), Analog RGB Output, Printout, Foot Switch Input |
| Power |
100- 240VAC±10%, 50/60Hz, 150VA |
| Specifications |
335mm (length) x 200mm (width) x 380mm (height): keyboard off 335mm (length) × 500mm (width) × 380mm (height): keyboard open |
| Weight |
Weight: about 11kg |
|