  • 1667050644378.jpg

Oocyte retrieval instrument negative pressure pump

The negative pressure pump ROCKET R29.655 of the oocyte retrieval instrument uses a safe and reliable diaphragm pump. The fluid valve is designed to prevent fluid substances from flowing back into the vacuum pump of the main engine. The air solenoid is simply controlled by feet, and the operation is safe.

Brand:

Model:


Product Information

The negative pressure pump ROCKET R29.655 of the oocyte retrieval instrument uses a safe and reliable diaphragm pump. The fluid valve is designed to prevent fluid substances from flowing back into the vacuum pump of the main engine. The air solenoid is simply controlled by feet, and the operation is safe.

Product use

The Rocket UK negative pressure aspirator has more than 20 years of history, used to provide smooth, low-volume vacuum under negative pressure, simple operation, low turbulence

Technical indicators

(1) Measurement range: 0-760mmHg

(2) Adjustable range: 0-400mmHg

(3) using vacuum diaphragm pump design

(4) the use of pneumatic control switch, free to open or close.

(5) With anti-overflow buffer bottle to protect the internal machine.

(6) The pressure is indicated by a pointer and marked with a green safety zone.

(7) With instantaneous pressure plugging of the needle, the sampling needle can be dredged by instantly raising the pressure.

Product Configuration

One vacuum negative pressure vacuum pump, two sets of one-time anti-overflow bottle suit, one foot switch, one power cord and one instruction manual.

Technical Data

Detailed Chinese operation guide, relevant information and certificate of instrument maintenance.

Get Quote

We will contact you within one working day. Please pay attention to incoming calls from your mobile phone.


Get Quote

Get Quote


We will be happy to provide you with quality products and services